Icon NMR logo

IconNMR Automation Run Status:
Day Experiments: 00:00
Busy Until: Wed 14:01
Night Experiments: 00:00

Setup List
Holder Status Name ExpNo Experiment User Time Title
1 Completed 180117-nmr 1 N sst.1H.lsh 1H Lineshape: CHCl3 in d6-Acetone nmr 00:01:46 1H Lineshape: CHCl3 in d6-Acetone Specification result of humpcal: Linewidth at 0.55% of signal height = ( < 6 Hz) Linewidth at 0.11% of signal height = ( < 12 Hz) Halfwidth = ( < 0.6 Hz)
2 Completed 180117-nmr 2 N sst.1H.sen 1H Sensitivity 0.1% Ethyl Benzene in CDCl3 nmr 00:01:04 1H Sensitivity 0.1% Ethyl Benzene in CDCl3 Specification result of sinocal: Best sino value found ( > 300 :1)
3 Completed 180117-nmr 3 N sst.13C.sen 13C Sensitivity 10% Ethylbenzene (EB) in d6-acetone nmr 00:05:06 13C Sensitivity 10% Ethylbenzene (EB) in CDCl3 Specification result of sinocal: Best sino value found ( > 200 :1)
4 Completed 180117-nmr 4 N sst.19F.qnp.sen 19F Sensitivity 0.05 Trifluourotoluene in CDCl3 nmr 00:00:47 19F Sensitivity 0.05 Trifluourotoluene in CDCl3 Specification result of sinocal: Best sino value found ( > 250 :1)
5 Completed 180117-nmr 5 N sst.31P.sen 31P Sensitivity 0.0485 M triphenylphosphate (TPP) in d6-acetone nmr 00:00:14 31P Sensitivity 0.0485 M triphenylphosphate (TPP) in d6-acetone Specification result of sinocal: Best sino value found ( > 150:1)
History List
Holder Time Name ExpNo Experiment User Load Atma Spin Lock Shim Acq Proc Title Remarks
5 12:18:24 180117-nmr 5 sst.31P.sen nmr + + + + + + + 31P Sensitivity 0.0485 M triphenylphosphate (TPP) in d6-acetone Specification result of sinocal: Best sino value found ( > 150:1) No Pdf file found in dataset: F:/data/nmr/nmr/180117-nmr/5/pdata/1/180117-nmr_5_1.pdf Check for printing system problems, Automation was stopped
4 12:09:06 180117-nmr 4 sst.19F.qnp.sen nmr + + + + + + + 19F Sensitivity 0.05 Trifluourotoluene in CDCl3 Specification result of sinocal: Best sino value found ( > 250 :1) No Pdf file found in dataset: F:/data/nmr/nmr/180117-nmr/4/pdata/1/180117-nmr_4_1.pdf Check for printing system problems
3 11:52:32 180117-nmr 3 sst.13C.sen nmr + + + + + + + 13C Sensitivity 10% Ethylbenzene (EB) in CDCl3 Specification result of sinocal: Best sino value found ( > 200 :1) No Pdf file found in dataset: F:/data/nmr/nmr/180117-nmr/3/pdata/1/180117-nmr_3_1.pdf Check for printing system problems
2 11:42:59 180117-nmr 2 sst.1H.sen nmr + + + + + + + 1H Sensitivity 0.1% Ethyl Benzene in CDCl3 Specification result of sinocal: Best sino value found ( > 300 :1) No Pdf file found in dataset: F:/data/nmr/nmr/180117-nmr/2/pdata/1/180117-nmr_2_1.pdf Check for printing system problems
1 11:05:33 180117-nmr 1 sst.1H.lsh nmr + + + + + + + 1H Lineshape: CHCl3 in d6-Acetone Specification result of humpcal: Linewidth at 0.55% of signal height = ( < 6 Hz) Linewidth at 0.11% of signal height = ( < 12 Hz) Halfwidth = ( < 0.6 Hz) No Pdf file found in dataset: F:/data/nmr/nmr/180117-nmr/1/pdata/1/180117-nmr_1_1.pdf Check for printing system problems
1 11:01:48 180117-nmr 1 sst.1H.lsh nmr x - - - - - - 1H Lineshape: CHCl3 in d6-Acetone Specification result of humpcal: Linewidth at 0.55% of signal height = ( < 6 Hz) Linewidth at 0.11% of signal height = ( < 12 Hz) Halfwidth = ( < 0.6 Hz) Sample Changer Fatal Error: Error: Error: STOP_CHANGER 82.2 SENSOR ERROR: SAMPLE AT MAGNET. LISH IS CLOSED !